Non conventional use
In case of non conventional use of the equipment danger for life and limb as well as danger for property
may occur. Therefore a non conventional use is prohibited. In case of non conventional use EMAG AG
will not be liable for possible damages and will not warrant for the correct and functionally work of the
components. In order to avoid non conventional use of the ultrasonic cleaning equipment the following
directions are to be considered:
• The equipment may only be operated with the voltage printed on the type plate on the backside of the
housing (shockproof socket).
• The equipment must not be used for cleaning of contact lenses.
• Before any kind of handling of the equipment (i.e. transport, movement etc.) it must be assured that the
power supply is disconnected and that the equipment chamber is at room temperature.
• No flammable or aggressive chemicals (i.e. acids etc.) must be used for cleaning.
• The equipment must only be operated in environments in which no explosive gases can occur.
• The equipment must only be operated indoor, in dry environment.
• In case of transportation damage, don `t connect the equipment with main source.
• The equipment must be protected against humidity and wetness – no water must be filled into the inter
or, or must intrude into the equipment.
Operation
If the equipment – after unpacking – is transported it must be assured that that the equipment is carried
in horizontal position and is hold below the bottom. Before connecting the equipment with the main
source it must be checked that the voltage of the main source is the same than the one stated on the
type label of the equipment (230V~ / 50 Hz). The equipment must only be installed on even areas in
order to allow air to circulate below the equipment. Attention must also be paid that the back side of the
equipment is at least 10 cm away from the wall or other obstacles in order to allow an adequate air
circulation. The power supply cable is to be connected with the equipment and thereafter with the main
source. Switch power supply switch to I (ON) and check whether the green control light is on.
Note: If the equipment is switched on the first time a light smell of heated ceramic fibre may
occur which may last for a short time. This is normal! Switch the time switch clockwise to the
wished time, max. 120 minutes. Adjust the temperature regulator to the wished temperature and
check whether the orange control light is on.
Note: The orange light may jitter a little bit until the adjusted temperature is reached. If the
equipment is switched on at room temperature it will take approx. 30 minutes until the adjusted
temperature is reached. The recommended operating temperature is +150°C to +180°C.
Shall the equipment operate for more than 120 minutes the time regulator is to be adjusted anti clockwise
to the position " ∞ ". In order to get a correct sterilisation the insert must not be overloaded. The load must
allow a good circulation of the hot air. When the adjusted temperature is achieved the sterilisation time
shall be between 30 and 120 minutes, depending on legal requirements and / or specifications of the
manufacturers of the items to be sterilised.
13