ERROR DESCRIPTION
Error
Wrong pH buffer solution or the
recognition of calibration solution out
of range
Press
not stable during calibration.
During calibration, the measuring value
is not stable for ≥3min.
Electrode zero electric potential out of
range (<-60mV or >60mV)
Electrode slope out of range (<85%or
>110%)
pH measuring range out of range
(<0.00 pH or >14.00pH) 1 Series
( <-2.00 pH or >16.00pH ) 5 Series