2 Use with factory setting
Installation
Before mounting and removing the sensor, make sure that no pressure
is applied to the system.
The unit is adaptable for various
process fittings by Aseptoflex
adapters (to be ordered sepa-
rately as accessories).
Mount adapter (B) to the sensor
first, then sensor + adapter to
the process connection by means
of a nut, a clamping flange or
similar (A).
If it is not possible to slide the fix-
ing element (A) down over the
top of the sensor: slide it up over
the bottom of the sensor before
the adapter is mounted.
NOTE: A guarantee for a long-term stable and maintenance-free fit-
ting, with no bug traps in the hygienic sealing of the metal seal
(Aseptoflex connection) is only valid for once-only mounting.
Mounting of the Aseptoflex
adapter
Step 1
Grease thread and sealing cham-
fer of the sensor and of the
adapter with the greasing paste
supplied.
The greasing paste is food-grade
(USDA-H1 84-201).
Make sure that the O-ring (C) is
correctly positioned.
14