PT2026 Installation and Safety Manual
 Always use a three-prong IEC detachable power cord that is grounded,
rated for 250 V – 10 A, and with adequate temperature ratings. The cord
should satisfy IEC 60245-1 for rubber insulated cables or IEC 60227-1 for
PVC insulated cables, as well as all applicable standards and legislation for
your region.
 To avoid electric shock, the power cord protective grounding conductor
must be connected to ground.
 The disconnecting device is the power plug. Ensure that it can be reached
in case of emergency.
 For continued fire protection, use only the specified fuse – see above.
 The PT2026 Main Unit contains ferromagnetic components and may be
subject to violent forces in a strong magnetic field gradient.
 This is a class A product. In a domestic environment this product may
cause radio interference in which case the user may be required to take
adequate measures.