Step/Range
Settings
17, Aux o/p
18, Aux o/p
19, Aux o/p
20, Aux o/p
21, Aux o/p
22, Aux o/p
XDL 56-4, XDL 56-4P, XDL 56-4T & XDL56-4TP
Step/Range
Settings
1, 56V 2A
0.01V 2A
2, 56V 2A
0.01V 2A
3, 56V 2A
4, 56V 2A
5, 56V 2A
6, 56V 2A
0.01V 2A
7, 56V 2A
8, 56V 2A
2V 0.001A
9, 56V 2A
2V 0.001A
10, 56V 500mA
2V 0.1mA
11, 56V 500mA
2V 0.1mA
12, 25V 4A
13, 25V 4A
14, 56V 500mA
2V 400mA
15, 56V 500mA
2V 400mA
16, 56V 500mA
2V 410mA
17, Aux o/p
18, Aux o/p
19, Aux o/p
20, Aux o/p
21, Aux o/p
22, Aux o/p
Adjust for:-
5V
5.00V ± 10mV
5V
5.00V on display
1V
1.00V on display
5V
5.00V ± 0.5mV
off
0.000 flashing 0.001
5V
Display equals load setting
or ammeter
Adjust for:-
10mV ± 0.5mV
0.010V on display
35V 2A
35V ± 0.5mV
35V 2A
35.000 on display
41V 2A
41.000 on display
10mV ± 0.5mV
35V 2A
35V ± 0.5mV
1mA ± 0.5mA
0.001 on display
0.1mA ± 0.05mA
flashing 0.0/0.1 on display
2V 4A
4A ± 0.5mA
2V 4A
4.000 on display
400mA ± 0.05mA
400.0 on display
410.0 on display
5V
5.00V ± 10mV
5V
5.00V on display
1V
1.00V on display
5V
5.00V ± 0.5mV
off
0.000 flashing 0.001
5V
Display equals load setting
or ammeter
Load
DVM
DVM
DVM
DVM
A readback zero
2.7A load, or
1.8Ω in series
with ammeter
Load
DVM
DVM
V readback zero
DVM
DVM
V readback span
DVM
V readback 13bit
DVM
DVM
milli-ammeter
milli-ammeter
A readback zero
milli-ammeter
o/p mA offset
milli-ammeter
mA readback zero
change load
ammeter
ammeter
A readback span
ammeter
o/p mA span
ammeter
mA readback span
ammeter
mA readback 13bit
DVM
DVM
DVM
DVM
A readback zero
2.7A load, or
1.8Ω in series
with ammeter
Detail
o/p V span
V readback
V readback
V readback
A readback
Detail
o/p 0V
o/p V span
o/p 0V
o/p V span
o/p A offset
o/p A span
o/p V span
V readback
V readback
V readback
A readback
41