- After sanitization shake the accessories vigorously and place them on a paper towel, or alternatively,
dry them with a jet of hot air (e.g. using a hair dryer).
- Dispose of the solution as recommended by the disinfectant manufacturer.
STERILIZATION
The accessories that can be sterilized are (C5.1-C5.2-C5.3-C5.4-C3-C4)
The sterilization procedure described in this paragraph is effective only if all the prescriptions herein
are met, all the components involved are previously sanitized and the procedure itself is validated
according to ISo 17665-1 rule.
Unit: Steam sterilizer with fractionated vacuum and overpressure according to EN 13060 rule.
Procedure: Package each component to be treated in sterile barrier system or packaging complying
with EN 11607 rule. Place the packed components in the steam sterilizer.
Run the sterilization cycle according to the instructions manual of the device, selecting a temperature
of 134° C and a duration of 10 minutes
Storage: Store sterilized components as per instructions of the sterile barrier package or system chosen.
AIR FILTER
The device has a suction filter (A3) that should be replaced when it is dirty
or changes color. Do not wash or reuse the same filter. Regular replacement
of the filter is required to help and ensure correct performance of the
compressor. Check the filter on a regular basis. Contact your local dealer or
authorized service center for spare filters.
Use original FLAEM accessories only.
AVAILABLE SPARE PARTS
Description
- RF6 PLUS nebuliser kit composed of:
nebuliser, mouthpiece with valve and nasal prong
- 2 m connection tube
- Mask Set composed of: adult mask, pediatric mask and elastic band
- Spare air filter kit (2 pcs)
SYMBOLS
Class II unit
Type-B unit
Important: check the operating
instructions
Alternate current
In compliance with European standard
EN 10993-1 Biological evaluation of medical
devices and 93/42/EEC Medical Devices Directive
TECHNICAL SPECIFICATIONS
Mod. FLAEM JUNIOR F2000
Voltage:
230V~ 50Hz 210VA
In conformity
Dir. 93/42/EEC
with:
(EN 60 601-1)
Approvals:
Max Pressure:
Compressor air output:
Noise level (at 1 m):
operation:
Dimensions
Weight
115V~ 60Hz 230VA
3.5 ± 0.5 bar
14 LPM approx
57 dB (A) approx
Continuous use
20(W)x18(D)x11(H) cm
1.8 Kg
Switch oFF
Switch oN
Do not use the unit while bathing
or showering
manufacturer
220V~ 60Hz 230VA
9
Code
ACo368P
ACo225P
ACo462P
ACo164P
100V~ 50/60Hz 230VA