The output signal
is too low.
The output signal
is too high
(> 1200 mV).
Spectrosense, Instructions for Use
• Check if the mirror is tightly screwed onto the
Spectrosense.
• Check if there are any foreign materials or air
bubbles in the optical path.
• Check the mirror surface. If it is scratched, replace
the mirror; order number 6.1250.010.
• Maybe a voltage adjustment of the Spectrosense
has to be done, see Section 5.2.
• Maybe a voltage adjustment of the Spectrosense
has to be done, see Section 5.2.
4 Maintenance