MAINTENANCE AND CLEANING
HANDPIECE
1.
After each use, remove and dispose of barrier sleeve.
2. Wipe the Radii Xpert handpiece with a disinfectant wipe
(benzalkonium base) or soft cloth moistened with isopropyl
alcohol/ethyl alcohol based cleaners.
3. Do not wipe electrical contacts at the base of the handpiece.
Use a dry cloth to clean if required.
4. Allow handpiece to air dry before placing a new barrier sleeve.
Note: If using anti-bacterial wipes, ensure they are not soaked in excess liquid.
Liquid residue must not enter any part of the handpiece or charger.
CHARGING DOCK
1.
Disconnect power from charging dock.
2. Wipe surface with a disinfectant wipe (benzalkonium base) or soft cloth
moistened with isopropyl alcohol/ethyl alcohol based cleaners.
3. Use an air gun to blow off any debris collected inside the handpiece holder
and carefully wipe clean without wetting the electrical contacts.
CLEANING AGENT APPROVED FOR USE
70% Isopropanol alcohol
Neutral-enzymic cleaner
Ethyl alcohol based
Antiseptic benzalkonium chloride wipes
Chlorhexidine gluconate
Glutaldehyde
STORAGE DURING EXTENDED PERIODS OF NON-USE
If the Radii Xpert is not to be used for an extended period of time, ie during holiday periods, fully charge the handpiece or keep it on the charging dock
with the power ON.
The Radii Xpert has a built in safety feature to prevent the battery from discharging totally.
DISPOSAL INSTRUCTION
Do not dispose unserviceable battery and curing lights as unsorted municipal waste. Hazardous substances are present in electronic
equipment and present potential risks to human health and the environment if disposed of in municipal landfills which are not designed
to prevent migration of substances into the soil and groundwater.
Please dispose of in accordance to your specific country regulations.
WARRANTY
SDI Ltd provides a three (3) year warranty against defects in materials or workmanship from the date of purchase to the original purchaser of this
product with the exception of the battery which has a two (2) year warranty. User is responsible for determining the suitability of the product for user's
application. If this unit is defective within the warranty period, SDI Ltd or your local approved service centre will repair the unit at no cost. This warranty
is only valid providing an inspection indicates that any such defect developed during normal and proper use subject to the conditions below.
SDI Ltd shall under no circumstances be liable for incidental or consequential damages. The warranty does not apply to any damage or defect caused
by the user's failure to follow the instructions for use or the other user error in operating or maintaining the unit. Although SDI Ltd offers services to
repair the Radii Xpert after the expiration of the warranty, SDI Ltd only guarantees the availability of such services and replacement parts for two (2)
years after the model has been discontinued.
Only the above warranty shall be applicable, all other warranties, expressed or implied are hereby excluded.
CLAIM
In case of a claim under warranty, the complete unit (handpiece, charging dock, AC power plug and LED attachment) is returned, carriage paid to the
dealer or directly to SDI Ltd together with a copy of the purchase document together with a clear description of the defect. Ensure you have your Radii
Xpert's serial number ready when communicating with the Customer Service team.
REGISTERING YOUR RADII XPERT WARRANTY
TO SPEED UP YOUR WARRANTY CLAIM, IT IS ADVISABLE TO REGISTER YOUR RADII XPERT WITH SDI LTD AS SOON AS POSSIBLE USING THE
FOLLOWING TWO OPTIONS:
1.
Fill out the warranty card included in kit and return with a copy of your receipt to: SDI Ltd PO Box 314 Bayswater, VIC 3153 AUSTRALIA
2. Alternatively goto http://www.sdi.com.au/warranty to register your warranty.
CAUTION
HIGH
INTENSITY
LIGHT
CLEANING AGENT NOT APPROVED FOR USE
Oxidising agents such as hydrogen peroxide
High alkaline detergents
Hand soaps and dish soaps
Abrasive cleaners
Acetone based cleaners
Bleach based cleaners
Any cleaners with phosphoric acid
Ammonia or iodine complex solution
Do not autoclave the LED light attachment
or any part of the Radii Xpert.
Liquids must not enter any part of the handpiece or
charger.
Do not spray liquids directly onto, or apply copious
amounts of liquids to the Radii Xpert. Instead, spray
solutions on a cloth to wipe.
Do not look directly into the light source.
The tip of the LED attachment is the only
emission aperture that relates to the
above hazard warnings
CAUTION: The light emitted may be harmful to the eyes.
Do not stare at the light source.
Spectral irradiance 47930Wm-2sr-1 - Conforms to IEC 62471
1500 mW/cm
2
(peak) - Avoid exposure -
Optical radiation is emitted from this aparture.
ENGLISH
7