12.
Verify that the two flanges are seated flush together. If necessary, slight-
ly rotate the Upper Microscope Assembly left and right until the exact fit
is achieved.
13.
Tighten the thumbscrew. The Upper Microscope Assembly is now
locked to the base portion of the microscope.
14.
Reinstall accessories to the beam splitter, unlock the operating micro-
scope and turn on the electrical power to the microscope and acces-
sories.
INSTALLING THE REINVERTER ON A MULTIPLE SCOPE UNIT
To install the ROLS Reinverter on a microscope containing a Surgeon's Scope and
Assistant Scope:
1.
Turn off the electrical power to the microscope and attached micro-
scope accessories.
2.
Lock the microscope in an easily accessible position.
3.
Remove any accessories attached to the Beam Splitter, including the
Observer's Tube.
4.
Remove the Surgeon's Binoculars, the beam splitter, and/or the laser filter.
Important
Do not remove the Assistant's Scope. For proper operation, the rein-
verter must be positioned below the Surgeon's Scope but above the
Assistant's Scope.
5.
Hold the ROLS Reinverter in a horizontal position with the Volk lettering
readable from the chief surgeon's position.
6.
Slide the lower Male Flange Lock on the reinverter into the upper
Female Flange Lock of the base portion of the microscope.
7.
Verify that the two flanges are seated flush together. If necessary, slight-
ly rotate the reinverter unit left and right until the exact fit is achieved.
8.
Tighten the thumbscrew on the microscope base. The ROLS unit is now
locked to the base portion of the microscope.
9.
Remove the Protective Plastic Cover from the reinverter.
10.
Slide the lower Male Flange Lock on the Surgeon's Scope Assembly
into the upper Female Flange Lock on the reinverter while observing the
assembly is in the correct orientation.
Important
For proper clearance, the laser filter or the beam splitter must be posi-
tioned between the reinverter and the Surgeon's binoculars.
11.
Verify that the two flanges are seated flush together. If necessary,
slightly rotate the Surgeon's Scope Assembly left and right until the
exact fit is achieved.
12.
Tighten the thumbscrew. The Upper Microscope Assembly is now
locked to the base portion of the microscope.
13.
Verify that all flanges are seated flush together. If necessary, slightly
rotate the appropriate assembly left and right until the exact fit is
achieved.
14.
Tighten all thumbscrews, reinstall accessories to the beam splitter,
unlock the operating microscope and turn on the electrical power to the
microscope and accessories.
USING THE REINVERTER
The ROLS Reinverter does not interfere with the microscope's functions.
IN/OUT INDICATOR
The In/Out Indicator of the ROLS Reinverter visually indicates when the prism is
introduced or removed from the optical pathway.
MAGNETIC LATCHING HANDLES
The Magnetic Latching Handles of the ROLS unit have been designed to be easily
removed, adjusted and sterilized by autoclaving or other methods.
Adjustment
Each handle may be independently adjusted to any angle by:
1.
Pulling the handle from the unit, rotating it to the desired angle, and
then reinstalling the handle on the splined end.
2.
Verify the edge of the handle and the edge of the In/Out Indicator are
seated flush together. If necessary, adjust each handle position slightly
until you hear a "click" sound, which will indicate the handle is in place.
Operation
When using a plano/concave lens or when the operating microscope is used in
other applications, use the reinverter in the Out position. To remove the prism from
the optical pathway, rotate either handle 60° until the In/Out Indicator reads Out.
When using any of the vitrectomy contact lenses, use the reinverter in the In posi-
tion To introduce the reinverter prism, rotate either handle 60° until the In/Out
Indicator reads In.