Calibration Of Axes; Calibrating The Artex Fixator; (Accessory) - IDC S1 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

5.6

Calibration of Axes

Following the set-up and connection to the PC, the
desktop scanner must be calibrated using the sup-
plied calibration model.
▷ Mount spacer plate No. 1, the fixation plate with
the clamp and the model plate (see
section 4.4.2).
Fig. 5
Calibration model
▷ Insert the calibration model
▷ Create a new patient case and start the scanner
software from the data base.
▷ Cancel the subsequent command prompt for
the 2D scan.
▷ Click on
Scanner Options > Axis calibration
start the axis calibration.
▷ Follow the software instructions.
If required:
▷ Re-enter the calibration model values already
entered during the installation by clicking on
Application > Register Calibration Object
The scanner should be recalibrated at reg-
ular intervals.
5.7

Calibrating the Artex fixator

(accessory)

The calibration of the fixator with the Spli-
tex calibration key should be checked at
regular intervals.
Before the Artex Fixator is used in the scanner, it
must be calibrated with the Splitex calibration key:
2
3
4
5
Fig. 6
1 Fixator, bottom element
2 Adhesive surface
to
3 Splitex plate
4 Splitex calibration key
5 Fixator, upper element
6 Lateral screw
7 Rear screw
▷ Open the Fixator.
.
▷ Place the Splitex calibration key 4 onto the
upper element 5.
▷ Place the Splitex plate 3 onto the calibration
key.
▷ Apply adhesive (Loctite) to bottom side of the
Splitex plate.
I N S T A L L A T I O N
1
7
6
E N
33

Publicité

Table des Matières
loading

Table des Matières