AMETEK Programmable Power Sorensen XDL Série Mode D'emploi page 8

Table des Matières

Publicité

Over-voltage Protection:
(OVP)
Over-current Protection:
(OCP)
Protection Functions:
METER SPECIFICATIONS
Display Type:
Voltage (CI mode):
Current (CV mode):
V x A:
ALARM OUTPUT
Isolated rear-panel open-collector output signal. User can select output to be activated for either
OVP, OCP, Overtemperature or Sense miswiring, or for any of those four faults.
KEYBOARD & ROTARY CONTROL
All functions, including the selection and set-up of the remote control interfaces, can be set from
the keyboard. The rotary jog control can be used to adjust output voltage and current settings in
a quasi-analogue mode.
INTERFACES (XDL 35-5P & XDL 56-4P only)
Full remote control facilities are available through the optional RS232, GPIB and USB interfaces.
Setting and readback resolutions are the same as for the Output and Meter specifications
respectively.
RS232:
IEEE-488:
USB:
Remote Command
Processing Time:
GENERAL
AC Input:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Cooling:
Store/Recall:
Safety:
EMC:
Size:
Weight:
6
Range 1V to 40V (XDL 35-5), 1V to 62V (XDL 56-4)
Resolution 0·1V; accuracy ± (2% + 0·5V)
Response time typically 100µs
Range 0·01A to 5·5A (XDL 35-5), 0·01A to 4·4A (XDL 56-4)
Resolution 0·01A; accuracy ± (0·2% + 0·01A)
Response time typically 35ms
Output trips off for OVP, OCP, over-temperature and Sense miswiring.
5-digit (Volts), 4-digit (Amps), 14mm (0·56") LED.
Resolution 10mV
Accuracy ± (0·1% of reading + 10mV)
Resolution 0·001A; 0·1mA on 500mA range
Accuracy ± (0·2% + 0·005A); ± (0·2% + 0·5mA) on 500mA range
Resolution 0·01W; 0·001W on 500mA range
Accuracy ± (0·3% + 0·05W); ± (0·3% + 0·005W) on 500mA range
Variable Baud rate, 19200 Baud maximum. 9-pin D-connector.
Single instrument or addressable RS232 chain operation.
Conforming with IEEE488.1 and IEEE488.2
Standard USB hardware connection.
Typically <80ms between receiving the command terminator for a
step voltage change at the instrument and the output voltage
beginning to change.
230V AC or 115V AC ± 10%, 50/60Hz. Installation Category II
250VA max.
+5ºC to +40ºC, 20% to 80% RH
−40ºC to + 70ºC
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Intelligent variable-speed fan.
Over-temperature trip shuts down output if internal temperatures
exceed predetermined thresholds.
Up to 10 set-ups can be saved and recalled via the keyboard or
remote interfaces.
Complies with EN61010−1
Complies with EN61326
140 x 160 x 290mm (WxHxD), excluding feet and terminals.
5·5kg

Publicité

Table des Matières
loading

Table des Matières