Metrohm KF 768 Mode D'emploi page 28

Four kf
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3.2.2.17
Info.Report
Output of the selected report.
3.2.2.18
Info.Report.Select
Selection of the report.
configuration: Configuration report. Not accessible when a determination is running.
parameters:
Parameter report. Only "live" parameters when a determination is running.
result:
Result report of the last completed determination.
The report output must be triggered with &Info.Report $G, see 3.2.2.17.
Reports which are sent automatically by the KF Oven start with a space (ASCII 32) and '.
This is followed by the individual identification for each report. Reports which are requested
via RS232 ($G) have the same individual identification of the report blocks, but are started
with ' only (no preceding space).
3.2.2.19
Info.Results.PurgeTime
Info.Results.CondTime
Info.Results.SampleHeatTime read only
Info.Results.LowTemp
Info.Results.HighTemp
Info.Results.GasFlow
Info.Results.LowFlow
Info.Results.HighFlow
Inquiry of the current results.
.PurgeTime: Purge time.
.CondTime:
Conditioning time.
.SampleHeatTime: Time during which the sample has been heated.
.LowTemp:
Lowest temperature during the sample heating.
.HighTemp:
Highest temperature during the sample heating.
.GasFlow:
Mean gas flow during the sample heating.
.LowFlow:
Lowest gas flow during the sample heating.
.HighFlow:
Highest gas flow during the sample heating.
3.2.2.20
Info.ActualInfo.Inputs.Status
Info.ActualInfo.Inputs.Change read only
Info.ActualInfo.Inputs.Clear
Info.ActualInfo.Outputs.Statusread only
Info.ActualInfo.Outputs.Changeread only
Info.ActualInfo.Outputs.Clear
Status sends the current status of the I/O lines. Change sends information regarding
whether a change in the status of a line has taken place since it was last cleared, Clear
clears the byte of Change.
For the output, the byte is converted to binary, e.g.
|0|0|0|0|1|0|1|0|
Line No.
|7|6|5|4|3|2|1|0|
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also pages 53, 54):
Inputs:
0 Start (Pin 21)
1 Stop (Pin 9)
2 Terminate (Pin 22) 2 Stop (Pin 4)
3 Pin 10
4 Pin 23
5 Pin 11
6 Pin 24
7 Cond.ok (Pin 12)
768 KF Oven
Output: 2 1 + 2 3 = "10"
Outputs:
0 Ready (Pin 5)
1 Start (Pin 18)
3 HeatSmpl (Pin 17)
4 Terminate (Pin 3)
5 Error (Pin 16)
6 Pin 1
7 Pin 2
3.2 Remote control commands
configuration, parameters, result
$G
$G
read only
read only
read only
read only
read only
read only
read only
read only
$G
25

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

01 série