RX mode
In RX mode, the pump must be connected to a RX sensor and through the proper configuration of the setup menu. The
device is responsible for ensuring the maintenance of the set-point value and the management of all the devices connected
to the pump terminal box.
In the configuration menu of the RX mode you can carry out the following settings:
Sensors made from glass or plastic material can be used, connect the probe as shown in fig. 10.
CI mode
In Cl mode, the pump must be connected to a Cl sensor such as a potentiometer and through the proper configuration of
the setup menu. The device is responsible for ensuring the maintenance of the set-point value and the management of all
the devices connected to the pump terminal box. The chlorine probe can be connected to the pump using the dedicated
cable: when the BNC is connected, the remaining two cables must be inserted into the cable sleeve and connected to the
"inputs" terminal of the pump (see Figure 11).
In the configuration menu of the Cl (ppm) mode you can carry out the following settings:
ppm mode
In ppm mode, the pump must be connected to a sensor such as a potentiometer in order to control certain chemicals.
By means of the proper configuration of the setup menu, the device is responsible for ensuring the maintenance of the set-
point value and the management of all the devices connected to the pump terminal box.
The probe for controlling certain chemical species can be connected to the pump using the dedicated cable: when the
BNC is connected, the remaining two cables must be inserted into the cable sleeve and connected to the ''inputs'' terminal of
the pump (see Figure 10).
In the Etatron list you can check the availability of the different types of sensors. The probes for chemical species are
connected to the pump as shown in fig. 11
In the configuration menu of the PPM mode you can carry out the following settings:
96
ENGLISH
Setpoint Configuration.
Probe Configuration.
Temperature setting (present if the MANUAL mode is enabled).
Current Output Setting
4A (-)
4B (+)
3
4
1
Setpoint Configuration.
Probe Configuration.
Temperature setting (present if the MANUAL mode is enabled).
Current Output Setting
Setpoint Configuration.
Probe Configuration.
Temperature setting (present if the MANUAL mode is enabled).
Current Output Setting
2
Fig. 11 – Connection Cl (ppm) and PPM probes
1. Pump power supply cable
2. CL probe
3. BNC Connector CL Probe
4. Power supply cable CL probe
Connect the power supply cables of the
probe to terminals 17 (4A- negative - white)
and 18 (4B – positive - brown) of the
terminal board.