Nivotemp NT MD
Possible errors
Problem /
Possible cause
Malfunction
Switching out-
• Switching out-
put not trigger-
put configured
ing when ex-
incorrectly
ceeding limits
• Switching out-
put defect
Switching out-
• Switching out-
put constantly
put configured
switching
incorrectly
• Switching out-
put defect
The analogue
• Wrong signal
doesn't receive
type set
the full/correct
output current
• Load too high
(current out-
put)
Analogue output
• Analogue out-
doesn't change
put configured
the output sig-
incorrectly
nal when the in-
put signal
changes
12
Bühler Technologies GmbH
Action
• In submenu Coux:
"Test Switching
Output" to ensure
normal mode
• In submenu Coux:
"Test Switching
Output" to test the
desired switching
output
• In submenu Coux:
"Test Switching
Output" to ensure
normal mode
• In submenu Coux:
"Test Switching
Output" to test the
desired switching
output
• In submenu Aoux:
Check and if neces-
sary set the correct
signal type (cur-
rent/voltage out-
put)
• Reduce load to per-
missible value
• In submenu CAnx:
„Test Analogue
Output" to ensure
normal mode
BX100022 ◦ 11/2020