Routine Maintenance; Repair; Ac Fuse Replacement; Cleaning - Covidien A-V Impulse 6060 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

To ensure that the A-V Impulse foot compression system remains in good operating condition,
it is important to keep it clean and carry out the routine maintenance procedures described
below.

14.1 Routine Maintenance

Interval
Routine Maintenance Procedure
Thoroughly clean external surfaces of the controller before transfer
As Required
between patients and after prolonged periods of storage. Attention
should be paid to the fan grills, ensure these are clean at all times.
Annually / 3000 hours 1. Inspect the AC power supply plug and cable for damage.
2. Perform the functional tests outlined in the Service Manual.

14.2 Repair

In the event of any problems occurring with this equipment, contact your local representative
for repair and the returns procedure.

14.3 AC Fuse Replacement

If the controller does not operate when AC power is applied check the fuses on the rear panel
and replace if necessary with fuses of the correct value and type; refer to the specification
section of this manual (Section 16.2). If the fuses blow again or the controller fails to power up
remove it from use and have it examined by authorized service personnel.

14.4 Cleaning

Warning: Disinfectants that are known to be corrosive to metals must not be used.
Use of Iodine can cause controller surface discoloration.
The controller enclosure can be cleaned with a soft cloth dampened with water or a mild
detergent. To sanitize the device, apply cleaning agents with a cloth or wipe. Avoid excessive
spraying, especially in the areas of the fan grills and connection ports on the back of the device.
If any liquid enters the grill or ports, then internal component damage will likely result. The
controller should be wiped with a clean, dry cloth afterward.
Do not immerse in any liquid. Do not use cleaning products containing ammonium chloride,
acetone or other aromatic solvents, as those chemicals will degrade the integrity of the case and
cause it to become embrittled and possibly crack.
The A-V Impulse foot compression system cannot be effectively sterilized by liquid immersion,
autoclaving, or ETO sterilization, as irreparable damage to the system will occur.
The table below provides recommended detergent information and their chemical components.
RECOMMENDED CLEANERS
Chemical component (with
approximate concentrations)
Dodecylbenzene Sulfonate,
Coconut Diethanolamide
diluted per instructions
0.5% bleach solution
EN-10
Commercial Example
Manu-klenz™*
Dispatch™*
A-V Impulse System

14.5 Disposal

The controller should be disposed of as outlined by local country regulations taking
environmental factors into consideration. Do not send the controller back to the manufacturer.

15. Warranty and Factory Service

Covidien warrants that your A-V Impulse foot compression system controller does not
contain defective material or workmanship. This warranty is for one year from the delivery
of the controller to the original purchaser. If within the first year there is a fault on the unit,
the controller should be returned to the Covidien Service Department at the address shown
below, and Covidien will free of charge, inspect the unit and replace any part which upon
their examination, appears to be defective provided that there is no evidence that the unit has
been tampered with, or mistreated. This warranty does not apply to the tubing assemblies or
the disposable Impad inflation pads or equipment damaged through shipping, tampering,
negligence, or misuse, including liquid immersion, autoclaving, or ETO sterilization.
If, within the first year of purchase the controller is serviced by personnel other than those
expressly authorized by Covidien, the warranty becomes void, and the company cannot be held
liable for any consequential damage caused as a result.
Some countries do not allow the exclusion or limitation of accidental or consequential damages,
so the foregoing limitation or exclusion regarding damages may not apply. Also, this Limited
Warranty gives you specific legal rights, and you may also have other rights, which vary, from
country to country.
The Service Manual is intended as a guide to technically qualified personnel when evaluating
instrument malfunctions, it is not to be construed as authorization to perform warranty repairs.
Unauthorized service will void the warranty.
If you have any service maintenance problems, contact Customer Services at Covidien.
If possible, the original packaging should be used to ensure safe arrival. Before shipping your
unit please call the number below to obtain a return material authorization number.
Covidien maintains a service facility with the capability to promptly repair an A-V Impulse foot
compression system controller. Controllers requiring repair must be shipped fully assembled,
prepaid and insured to:
CANADA Covidien Canada
7500 Trans Canada Highway
Pointe-Claire, Qc H9R 5H8
877-644-8926
AUSTRALIA Australian Service Centre
52A Huntingwood Drive
(Corner Liberty Road)
Huntingwood NSW 2148
Toll Free: 1800 350 702
Tel: (02) 9678 2256
Fax: (02) 9671 8118
A-V Impulse System
UNITED
Covidien
STATES
5920 Longbow Drive
Boulder, CO 80301
1- (800) 255-8522
EN-11