Operation
Cleaning
Clean every-day soiling using a moist cloth and a mild household cleaning agent. Do not use an
aggressive cleaning agent or solvent to clean the instrument.
Definition of the over-voltage categories
Measurement category II: Measurements on the power circuits fitted with a direct plug connection
to the low voltage system. Normal short circuit current < 10 kA.
Measurement category III: Measurements within the building installation (stationary consumers
with non-plug-in connection, distributor connection, permanently installed equipment in the distrib-
utor).
Normal short circuit current < 50 kA.
Measurement category IV: Measurements at the source of the low voltage installation (meters,
mains connection, primary over-current protection). Normal short circuit current >> 50 kA.
When establishing the measurement category for a combination of test probe assembly and
measuring instrument, the lowest category, either of the test probe assembly or the measuring
instrument, always applies.
Technical data
Display
Power supply for the transmitter
Power supply for the receiver
Dimensions
Weight
Operating temperature
External voltage protection
Degree of protection
18
Optical and acoustic
2 x 1.5V type AA
3 x 1.5V type AA
245 x 60 x 29 mm (receiver)
135 x 60 x 33 mm (transmitter)
318 g (inc. batteries)
5 °C to 40 °C
to 400 V
IP 31
Testboy
®
TB 27 Analog