Electrical connection
2)
e.g. HBM‐extension cable, 6 wires:
KAB8/00‐2/2/2 (sold by the meter, order no. 4‐3301.0071 = grey or 4‐3301.0082 = blue)
CABA1 (cable reel, order no. 1‐CABA1/20 = 20 m or 1‐CABA1/100 = 100 m in length), grey
CABE2 (cable reel, order no. 1‐CABE2/20 = 20 m or 1‐CABE2/100 = 100 m in length), blue
14
Wiring code 4‐wire circuit
Only applies to versions with 4‐wire connection
cable.
(black)
(white)
(green)
(red)
(yellow)
Cable extension
For extending cables, only use shielded low capacity
measurement cables (measurement cables provided by
2)
HBM
meet these requirements). Ensure that there is a
perfect connection with the lowest possible transition
resistance. If you use the 6‐wire circuit, you may neglect
any interference by a change in the resistance of the
extension cables. However, if you extend the cable using
the 4‐wire technique, the characteristic value deviation
can be removed by adjustment. Temperature influences
will not be compensated.
A1130-5.0 HBM: public
Excitation (-)
Signal (+)
Excitation (+)
Signal (-)
Shield, connected with housing
HLC / BLC / ELC