Trouble shooting
Error message
Problem
Autotest in progress, device not yeat ready for.
Large fluctuation in ambient temperature.
Ambient temperature below 10°C or above 40°C
(<50°F, >104°F)
Device no longer functioning correctly.
(1) Forehead thermometer mode: the temperature
measured is higher than 42,2°C (108°F).
(2) SCAN mode: the temperature measured is higher
than 80°C (176°F).
(1) Mode thermomètre frontal: la température
mesurée est inférieure à 34°C (93,2F).
(2) Mode SCAN: la température mesurée est
inférieure à -22°C (-7,6°F).
Autotest not successful.
Solution
Wait until the forehead symbol no longer flashes.
Store the device for at least 30 minutes in the room where the measure-
ment will be taken. The ambient temperature must be between 10°C and
40°C (50°F, 104°F).
Take out battery for approx. 1 minute and insert again. If the error
message is repeated, consult your dealer or customer service.
Operate the thermometer only between the specified temperature
ranges. If necessary, clean the sensor tip. In the event of a repeated error
message, contact your dealer or customer service.
Operate the thermometer only between the specified temperature
ranges. If necessary, clean the sensor tip. In the event of a repeated error
message, contact your dealer or customer service.
Insert new battery.
38