The best way to differentiate the Generation 2 pump from the Generation 1 pump is the inline filter that is visible on the
Generation 2 pump on the left-hand side. The product labels are also different.
Generation 1 Pump
Part number 116885 (yellow) and 118933 (black)
Operation Information for the Generation 1
pump
Using the sintered filter
For temperatures -10°C to +0°C (14°F to 32°F)
The maximum tube length when using the metallic sin-
tered filter is 10 ft. (3 m).
For temperatures 0°C to +50°C (32°F to 122°F)
The maximum tube length when using the metallic sin-
tered filter is 30 ft. (9.1 m).
Using the sample probe
For temperatures -10°C to +0°C (14°F to 32°F)
The maximum tube length when using the sample probe
is 10 ft. (3 m).
a Warning
When using the sample probe at -10°C to +0°C
(14°F to 32°F), keep the sample probe in your
hand .
For temperatures 0°C to +50°C (32°F to 122°F)
The maximum tube length when using the sample probe
is 10 ft. (3 m).
ERRATA CARD
Inline filter
Part number 130916 (yellow) and 130917 (black)
Operation Information for the Generation 2
pump
Using the sintered filter
For temperatures -20°C to +50°C (-4°F to 122°F)
The maximum tube length when using the metallic sin-
tered filter is 67 ft. (20 m).
Using the sample probe
For temperatures -20°C to +50°C (-4°F to 122°F)
The maximum tube length when using the sample probe
is 10 ft. (3 m)
see over for more important pump
GasAlertMicro 5 Series
Generation 2 Pump
information
131220