MICROPUMP 75211-50 Mode D'emploi page 15

Table des Matières

Publicité

SPECIFICATIONS
Output Speed:
40 to 3600 r/min
Operating Temperature:
0˚ to 40˚C (32˚ to 104˚F)
Storage Temperature:
–25˚ to 65˚C (–13˚ to 149˚F)
Chemical Resistance:
ABS plastic case with a polyester label,
gold, irridite-coated aluminum chassis
and zinc-plated screws.
All materials withstand standard
cleaning solvents.
Line Voltage Limits:
90–130 V or 190–260 V, 50–60 Hz
Power Output:
0.1 hp (0.075 kW)
Maximum Current:
115 V Units:
Controller: 2.2A, shorted
output conditions
230 V Units:
1.1A, shorted output conditions
Start/Stop–Local Control:
Input Voltage:
Function Disable:
15V DC Typ.
Function Enable:
0.8V DC Max.
Input Current:
Function Disable:
100 µA Max. leakage
Function Enable:
1.5mA Max.
Maximum Torque:
30 oz-in (2 kg•cm)
Speed Regulation:
Line:
±1%
Load:
±2%
Speed Drift:
±10%
Enclosure Rating:
IP23 per IEC529
Humidity (non-condensing):
10% to 90%
Altitude:
Less than 2000 m
Display:
Green LED
Dimensions (L
W
H):
11-1/2 in
(292 mm
Weight:
11 pounds (5 kg)
Remote Control
4–20 mA Input:
250 ohms typical input impedance
referenced to signal ground.
4 mA, (Stop); 20 mA, (Full Speed)
Accuracy: 3% F.S., Linear Resolution
Overload Capability: 10V or 40 mA max.
2–10 V Input:
10 K ohms typical input impedance
referenced to signal ground.
2 V, (Stop); 10 V, (Full Speed)
Accuracy: 3% F.S., Linear Resolution
Overload Capability: 15V max.
Compliance:
115V
115V: UL508, CSA C22.2, No. 14-M91
230V
230V (For CE Mark):
EN61010-1/A2: 1995
(EU Low Voltage Directive) and
EN61326-1/A1: 1998
(EU EMC Directive)
Pollution Degree:
Pollution Degree 2 per IEC 664
(Indoor use—lab, office)
Installation Category:
Installation Category II per IEC 664
(Local level—appliances,
portable equipment, etc.)
8 in
7-1/4 in
203 mm
184 mm)
Cole
REPLACEMENT PARTS
AND ACCESSORIES;
SPECIFICATIONS
ACCESSOIRES ET PIÈCES DE
RECHANGE ;
CARACTÉRISTIQUES
TECHNIQUES
ERSATZTEILE UND ZUBEHÖR;
TECHNISCHE DATEN
PIEZAS DE REPUESTO Y
ACCESORIOS;
ESPECIFICACIONES
ACCESSORI E RICAMBI;
SPECIFICHE TECNICHE
®
Parmer
15

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

75211-55

Table des Matières