External Linearization - Sartorius Signum Notice D'installation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

External Linearization

Notes on Settings
!
External linearization can be performed
on verified weighing instruments only
when the menu access switch is open.
– The "external linearization" function
must be assigned to the J key
(menu item 1.9.6 or 1.9.7.).
Procedure
(
J
)
Following external linearization,
!
close the menu access switch and reset
the ) key (> 2 sec) to its previous
function (e.g., external calibration/
adjustment with user-defined weights)
under menu item 1.9.
Zero the weighing platform
Start linearization
After approx. 2 seconds, you are prompted to place the first linearization weight
on the scale.
Place the prompted weight on the scale. After a brief pause, the difference between the
measured value and the true mass is displayed.
Store the linearization weight (press ( to cancel).
You are prompted to place the second linearization weight on the scale.
Repeat the steps described above for all linearization weights.
After you store the last linearization weight, you are prompted to unload the weighing
instrument.
Unload the weighing instrument. After a brief pause, the zero point is stored
automatically and the Signum display and control unit automatically returns to the
normal weighing mode.
17 17

Publicité

Table des Matières
loading

Table des Matières