Selecting The Configuration Mode - Sartorius Signum Notice D'installation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

)
k k )
k
)
k
repeatedly
ADC-CON
)
STAND.
SAVE
12

Selecting the Configuration Mode

Open the menu access switch:
§ Remove the cap that covers the menu access switch on the left-hand side of the back
of the display and control unit housing
§ Move the switch to the left ("menu accessible" position)
Activate the service mode (see previous page)
Confirm Setup menu: press )
Select the "Reference scale" menu item (COMSPEC) and confirm: press k twice,
and then press )
Select the "Analog/digital converter" menu item (ADU-232): press k
Confirm the ADU-232 menu item: press )
Select the "ADC configuration" menu item (ADC-CON): press k repeatedly
Confirm the ADC-CON menu item: press )
Select the "Standard" (STAND.) or "Verified" (VERIF.) configuration mode.
In this example, "Standard" is selected.
Subsequent configuration steps are illustrated in detail on the next two pages.
Following configuration, select the SAVE menu item to save your settings.
The A/D converter, in conjunction with the weighing platform, can now be used like any
standard weighing platform.
Close the menu access switch
Following ADC configuration, calibration/adjustment and linearization must be performed
on the weighing platform (see page 18, "Calibration and Adjustment," and page 20,
"Calibration and Adjustment without Weights" for details).

Publicité

Table des Matières
loading

Table des Matières