Elektrische Installation • Instalación eléctrica
Installazione elettrica • Instalação eléctrica
C
H
K

A
B
E
F
D
VAC - voltages
2.2
IN1
ext. Test
2.3
0V
0V
0.8
normally open
0.7
common
normally closed
0.6
0.2
PE
Earth/Ground
0.1
N
Neutral
L
Phase
0.0
 VDC - voltage
2.2
IN1
ext. Test
*)
2.3
0V
0V
0.8
normally open
0.7
common
0.6
normally closed
±24V
+24 VDC (0V)
±24V
0V (+24 VDC)
english
• Lift off domed cover (A) after removing
the 4 screws (B).
• Take power supply board (C) out of the
domed cover (A) after removing t h e
screw (D).
• Guide cables for power supply (E)
and potential-free contact (F) through
screwed cable fittings.
• Terminals
Check type plate (G) for permissible
mains voltage and ensure confor-
mity!
VAC power supply
24 VDC power supply +24 VDC (0V)
In the case of 24 VDC operation, do
not connect + VDC to frame because
the internal housing potential of the
device is negative.
• Connect potential-free contact (F) to
terminals 0.6 - 0.7 (fail safe) or 0.7 - 0.8
(open during malfunction).
• Pull cable (E + F) tight and screw down
cable fittings.
• Screw power supply board (G) with
screw (D) into domed cover (A)
• Plug ribbon cable (H) into control PCB
(K)
• If the individual wires have not been
*)
screwed into the plug, they should be
attached as follows:
1.0 = brown
1.1 = blue
2.0 = black
• Put on top of cover (A) and tighten the
4 screws (B)
External Test
see page 14-15
*) 0...1 V Test activ
5...36 V Test inactiv
Please note:
The power supply board (C) is in a
reverse position (upside down) in the
domed cover (A).
During no-load operation, a voltage of up
to 36 VDC may be measured at terminals
1.0 and 1.1 (plug ribbon cable (H)).
Please ensure that the installation is
carried out according to the valid regu-
lations.
0.0 L
0.1 N
0.2 PE
0V (+24 VDC)
BEKOMAT 16 CV