7. CHANGING BETWEEN CELSIUS AND FAHRENHEIT
This thermometer can display temperature measurements in either Fahrenheit or Celsius. To switch the display between °C and °F, simply turn
OFF the unit, press and hold the START button ②. This thermometer can display temperature measurements in either Fahrenheit or Celsius.
To switch the display between °C and °F, simply turn OFF the unit, press and hold the START button ② for 5 seconds; after 5 seconds, the
current measurement scale («°C» or «°F» icon) will be flashing on the display ⑫. Change the measurement scale between °C and °F by pressing
the START button ②. When the measurement scale has been chosen, wait for 5 seconds and the unit will automatically enter the «ready for
measuring» mode.
8. HOW TO RECALL 12 READINGS IN MEMORY MODE
This thermometer can recall the last 12 readings.
• Recall mode ⑬ : Press the START button ② to enter Recall mode when power is off. The memory icon «M» will flash.
• Reading 1 - the last reading ⑭ : Press and release the START button ② o recall the last reading. Display 1 alone with memory icon.
• Reading 12 - readings in succession : Press and release the START button ② consecutively to recall the last 12 readings in succession.
Pressing and releasing the START button ② after the last 12 readings have been recalled will resume the above sequence from reading 1.
9. ERROR MESSAGES
• Measured temperature too high ⑮ : Displays «H» when measured temperature is higher than 100.0°C (212.0°F) in the ear mode or 42.2°C
(108.0°F) in the forehead mode.
• Measured temperature too low ⑯ : Displays «L» when measured temperature is lower than 0°C (32.0°F) in the ear mode or 34.0°C (93.2°F)
in the forehead mode.
• Ambient temperature too high ⑰ : Displays «H» in conjunction with the «
» when ambient temperature is higher than 40.0°C (104.0°F).
• Ambient temperature too low ⑱ : Display «L» in conjunction with the «
» when ambient temperature is lower than 5.0°C (41.0°F) in the ear
mode or 16°C (60.8°F) in the forehead mode.
• Error function display ⑲ : The system has a malfunction.
• Blank display ⑳ : Check if the battery has been loaded correctly. Also check polarity (<+> and <->) of the battery.
• Flat battery indication ㉑ : If the steady battery icon is the only symbol shown on the display, the battery should be replaced immediately.
10. CLEANING AND DISINFECTING
Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the thermometer casing and the measuring probe. Ensure
that no liquid enters the interior of the thermometer. Never use abrasive cleaning agents, thin-ners or benzene for cleaning and never immerse the
instru-ment in water or other cleaning liquids. Take care not to scratch the surface of the probe lens and the display.
11. BATTERY REPLACEMENT
This instrument is supplied with one lithium battery, type CR2032. Replace with a new CR2032 battery when the flashing battery symbol appears
on the display ㉑. Remove the battery cover by sliding it in the direction shown. Remove the battery and replace with a new one ㉒.
Batteries and electronic instruments must be disposed of in accordance with the locally applicable regulations,
not with domestic waste.
12. TECHNICAL SPECIFICATIONS
Measurement range : Ear Mode: 0-100.0 °C (32-212.0 °F). Forehead Mode: 34-42.2 °C (93.2-108.0 °F).
Resolution : 0.1 °C / °F.
Measurement accuracy : Laboratory : ±0.2 °C, 32.0 - 42.2 °C. (±0.4 °F, 89.6 - 108.0 °F).
Display : Liquid Crystal Display, 4 digits plus special icons.
Acoustic :
• The unit is turned ON and ready for the measurement: 1 short beep.
• Complete the measurement: 1 long beep (1 sec.) if the reading is less than 37.5°C (99.4°F), 10 short «beep» sounds, if the reading is equal
to or greater than 37.5°C (99.4°F).
• System error or malfunction: 3 short «bi» sounds.
Memory :
• Auto-Display the last measured temperature.
• 12 readings recall in the Memory Mode.
Backlight :
• The display light will be GREEN for 4 seconds, when the unit is turned ON.
• The display light will be GREEN for 5 seconds, when a measurement is completed with a reading less than 37.5°C (99.4°F).
• The display light will be RED for 5seconds, when a measurement is completed with a reading equal to or higher than 37.5°C (99.4°F).
Operating temperature : Ear Mode: 5-40.0 °C (41-104.0 °F). Forehead Mode: 16-40.0 °C (60.8-104.0 °F); 15 - 95 % relative maximum humidity
Storage temperature : -25°C to +55°C (-13°F to 131°F); 15 - 95 % relative maximum humidity
Automatic Switch-off : Approx. 1 minute after last measurement has been taken.
Battery : CR2032 Battery (X1) 3V - at least 1000 measurements.
Dimensions : 107 x 50 x 34 mm.
Weight : 55g (with battery), 52g (w/o battery).
13. WARRANTY
This product is warranted for a two-year period from the purchase date. However, this warranty does not apply to damages, in particular, resulting
from non-compliance with this guide, incorrect manipulations, shocks or inappropriate installation of the battery. It does not apply either to the
battery or the package.
This device complies with the requirements of the Medical Device Directive 93/42/EEC. Technical alterations reserved.