Any malfunctions during operation will be identified by an error message on the display (LED (P)).
Proceed as follows in such cases:
- Switch off device using the main switch (E)
- Carry out corrective measures
- Restart device
Error code
Cause
E 3
Temperature inside device too high
E 6
Break in safety circuit
E 25
Heating and switching element monitoring
E 26
Difference between temperature of safety
sensor and temperature of control sensor
control temperature > (safety temperature
+ 40 K))
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA service department,
- send the device for repair, including a short description of the fault.
Error codes
Effect
Solution
Heating off
Switch off device and allow to cool down
Heating off
- Plug in contact plug (D)
- Plug in PT 1000 contact thermometer /
- Replace faulty connecting cable, plug, or
Heating off
- Switch off device
- Warning! Only to be carried out by autho-
Heating off
- Switch off device
- Warning! Only to be carried out by autho-
temperature sensor
contact thermometer
rised service personnel: Carry out an internal
test on the device to check the plug-in
connector for the heating element
rised service personnel: Carry out an internal
test on the device to check the temperature
sensor
15