Error Codes - IKA Dry Block Heater 1 Mode D'emploi

Masquer les pouces Voir aussi pour Dry Block Heater 1:
Table des Matières

Publicité

Any malfunctions during operation will be identified by an error message on the display (LED (Q)).
Proceed as follows in such cases:
 Switch off device using the main switch (E)
 Carry out corrective measures
 Restart device
Error code
Cause
E 3
Temperature inside device too high
E 6
Break in safety circuit
E 25
Heating and switching element monitoring
E 26
Difference between temperature of safety
sensor and temperature of control sensor
control temperature > (safety temperature
+ 40 K))
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA
®
service department,
- send the device for repair, including a short description of the fault.

Error codes

Effect
Heating off
Heating off
Heating off
Heating off
Solution
Switch off device and allow to cool down
- Plug in contact plug (D)
- Plug in PT 1000 contact thermometer /
temperature sensor
- Replace faulty connecting cable, plug, or
contact thermometer
- Switch off device
- Warning! Only to be carried out by autho-
rised service personnel: Carry out an internal
test on the device to check the plug-in
connector for the heating element
- Switch off device
- Warning! Only to be carried out by autho-
rised service personnel: Carry out an internal
test on the device to check the temperature
sensor
15

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Dry block heater 2Dry block heater 3Dry block heater 4