Product specifications
Category
Product
Product size
Weight or packing unit
Maximum temperature
Voltage, frequency,
power consumption
Type of protection against
electric shock
Degree of protection against
electric shock
Degree of protection against
harmful ingress of water
Mode of operation
※ The BIO-BELT will automatically turn off after 2 hours.
Class II ME equipment (single fault safe)
Touch Current
Ground Leakage Current
Testing in accordance with IEC 60601-1-2 indicates the following:
The BIO-BELT needs no special precautions regarding EMC and needs to be installed and put into service.
Portable and mobile RF communications equipment have no adverse effects on the BIO-BELT.
13
BIO-BELT
210 x 460 mm
1) main body of the BIO-BELT containing semi-precious stones:
Length 46 cm x width 21 cm
2) Total length including the belt: 140cm (Regular) or 160cm (XL)
1) Controller: 0.38 KG
2) Main body of the belt: 0.93KG
3) Overall weight: 1.6 KG
149°F / 65°C (The BIO-BELT is programmed to prevent overheating.)
120V~, 60Hz, 50W
Operating temperature : 95 ~ 149°F / 35 ~ 65°C
Variance: +/- 9°F / 5°C
Class II ME equipment
Class BF applied parts
- Applied parts: Top of Mat
Ordinary equipment (IPX0)
- Equipment not suitable for use in presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
Continuous operation
Normal Mode
< 100 microA
< 5 mA
Contents
Single Fault Mode
< 500 microA
< 10 mA