residue adhering to the light guide. If this is the case, immediately remove the residues and check that the surface
of the light guide has not been damaged. If damage is visible, replace the light guide as the lamp's power could be
significantly reduced.
- Under normal conditions of use, the lamp's power does not vary if the battery is properly charged. Consequently, it
is not necessary to check power as with ordinary polymerization lamps. Nevertheless, in case of doubt, check its
power using your light guide.
- The battery located at the bottom of the handpiece can be replaced by unscrewing it.
- After removing the light guide, check that the LED is clean and undamaged. If necessary, clean it using a medical-
quality dry air jet (free from compressor oil).
- During verification of the power level of the handpiece, if the translucent handpiece holder base is red, check if
the LED, light guide and verification window are clean.
- In the event of dust being present, clean using a dry air jet.
- If the problem persists or if the light guide or the verification window are damaged, the lamp must be returned to
the after-sales service.
In the event of an anomaly, contact the supplier of the unit rather than just any repairer, which may return
your unit in a hazardous state for both you and your patients.
The technical service of your supplier is at your disposal for any technical problems encountered on your unit.