2.3 Précision de mesure
Every electromagnetic flowmeter is calibrated by direct volume comparison. The wet calibration
validates the performance of the flowmeter under reference conditions against accuracy limits.
The accuracy limits of electromagnetic flowmeters are typically the result of the combined effect
of linearity, zero point stability and calibration uncertainty.
Reference conditions
• Medium: water
• Temperature: +5...+35°C / +41...+95°F
• Operating pressure: 0,1...5 barg / 1,5...72,5 psig
• Inlet section: ≥ 5 DN
• Outlet section: ≥ 2 DN
X [m/s]: flow velocity
Y [%]: deviation from the actual measured value (MV)
Précision
Version compacte avec IFC 300
DN15...100 / ½...4¨
DN150...300 / 6...12"
Version compacte avec IFC 100
DN15...300 / ½...12"
Version compacte avec IFC 050
DN15...300 / ½...12"
En option pour IFC 050 et IFC 100; étalonnage étendu en 2 points pour une précision optimisée.
Pour plus de détails sur la précision optimisée, voir la documentation du convertisseur de
mesure.
09/2019 - 4000631405 - TD OPTIFLUX 5000 FL R10 fr
CARACTÉRISTIQUES TECHNIQUES
Précision
± 0,15% de la vm + 1 mm/s
± 0,2% de la vm + 1 mm/s
Précision
± 0,3% de la vm + 1 mm/s
Précision
± 0,5% de la vm + 1 mm/s