Chauvin Arnoux E6N Notice De Fonctionnement page 18

Masquer les pouces Voir aussi pour E6N:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
English
Switch
The switch has 3 positions:
OFF : Of position
Range 10 mV/A
Range 1 V/A
NB: This range increases the sensitivity of the clamp whilst reducing the measurement
extent.
Battery indicator
This LED is unlit when the switch is «OFF». In normal use, the green LED is lit.
When the switch is no longer in the OFF position and the LED is unlit, this warns that the
value of the battery voltage is too low or that the cell is no longer supplied with power, thus
showing that there is a fault in the clamp.
Resetting zero on the clamp
The thumbwheel makes it possible to reset the output voltage to zero. It is thus possible to
overcome the different errors due to thermal shifts, the Earth's magnetic field, the
environment, and residual induction.
Can be turned 20 times ± 1 turn.
Battery compartment cover
This compartment is in the lower handle, the cover is held in place by a tool release screw.
Lead and plugs
Dual insulation twin core lead, untwisted 1.50m long terminated by 2 male safety plugs
(dia 4 mm).
Arrow
Arrow to show the direction of the primary current.
18

Publicité

Table des Matières
loading

Table des Matières