TROUBLESHOOTING
Type of problem
Control lamps do not illuminate when
the unit is switched on
No liquid flow when the Foot pedal is
activated
No ultrasonic oscillations when the Foot
pedal is activated
Decreasing or insufficient ultrasonic
power
Water leakage between handpiece and
handpiece hose
Water leakage
52
Solutions
•
Check all the connections of the unit
•
Switch on the unit with the ON/OFF button
•
Check the power supply of the dental office
•
Send the unit to an approved EMS repair center
•
Turn the liquid flow rate knob
•
Check the connections between the handpiece hose and the handpiece
•
Check the connections of the unit
•
Check the connection of the Foot pedal
•
Check the filter
•
Disconnect the handpiece and used with compressed air (max. 3 bar) to prevent blockage
•
If yes, send the handpiece to your EMS service team
•
Send the unit to an approved EMS repair center
•
Switch on the unit with the ON/OFF button
•
Check the connections between the handpiece hose and the handpiece
•
Check the connections of the unit
•
Check the connection of the Foot pedal
•
Check that the instrument is screwed on correctly
•
Increase the power of the ultrasounds to the maximum
•
Send the unit to an approved EMS repair center
•
Increase the power of the ultrasounds to the maximum
•
Check that the instrument is screwed on correctly
•
Check the wear of the instrument and replace it if necessary
•
Send the handpiece to an approved EMS repair center
•
Check the O-ring of the handpiece
•
Send the handpiece to an approved EMS repair center
•
Check the water hose connector
•
Send the unit to an approved EMS repair center