Dreve PCU LED N2 Mode D'emploi page 18

Photopolymérisateur pour le processus l'impression cao / fao en 3d
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

PCU LED N
2
8.4. Operation: Process level
34
34
Access to the
configuration level
Switching on the power switch
at the rear of the device and
simultaneously pressing the
control button takes you to the
process level. (See chapter 8.1)
Entry of the access code 217
(Status 03 / 2017)
Please confirm the code after
entry. The access code cannot
be freely selected and can also
change in the case of a firm-
ware update.
Configuration level
in overview
1
1 – Orientation line:
A vertically oriented narrow bar at
the left and the right margin of the
selected window is provided for
guidance purposes
Selection of the process
parameters
10 programme blocks avail-
able for programming
Green – activated for the
user level
White – not activated for the
user level
Setting the process para-
meters – press the rotary
button > 5 seconds
2
3
5
4.1
4.2
1
Setting the process
parameters
1 – Return button
2 – Programme name
(freely definable by turning
the rotary button)
3 – Intensity setting for LED sources
(freely definable by turning the
rotary button)
4 – Exposure medium
4.1 Green – nitrogen active
4.2 Green – vacuum active
4 .1 and 4.2 not marked – air active
5 – Exposure time
4
Setting the LED intensities
The intensity can be shown in
percent per panel. There are
3 LED spots on each panel.
Attention: If the predeter-
mined values are modified for
the appropriate material, the
final material properties indi-
cated by the manufacturer can
be changed.
Checking the vacuum function
The vacuum function is
checked on the basis of a
special vacuum value being
reached after a defined time.
The result is shown on this
screen after the test.
35
35

Publicité

Table des Matières

Dépannage

loading

Table des Matières