Operation Mode - Industrial Scientific TANGO TX1 Guide De Référence

Masquer les pouces Voir aussi pour TANGO TX1:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 21
Operation Basics

Operation mode

If enabled, these tasks can be completed – by qualified personnel – from operation mode:
The time-weighted average (TWA) and the short-term exposure limit (STEL) readings can be viewed and cleared.
The unit can be zeroed or zeroed then calibrated.
The unit can be bump tested.
The peak reading can be viewed and cleared.
The buttons are used as follows:
Press
to scroll through the operation-mode loop (shown below).
Press
to initiate a task or to clear a reading.
When a unit is in alarm, a long press on
When
and
are simultaneously pressed and held for three seconds, the unit will complete a self-test.
When no button is pressed for 30 seconds, the gas-monitoring screen is activated.
Operation-mode loop
Gas monitoring
Calibration date
Peak reading
TWA reading*
*Activated only when enabled for operation-mode access.
4
will reset a latched alarm; it does not disable an enabled latch.
Initiate zero*
STEL reading*
Initiate bump test*

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

17154831

Table des Matières