specifications
6
Size
Weight
Electrical Requirements
Sound Level
Power Consumption
Maximum Compressor Pressure
Neb. Operating Pressure
Compressor Free Air Flow
Operating Temperature Range
Operating Humidity
Storage/Transport Temperature Range -40°to +158°F (-40° to +70°C)
Storage/Transport Humidity
Supplied Nebulizer Capacity
Supplied Nebulizer Nebulization Rate
Supplied Nebulizer MMAD
18070 Equipment Classifications:
Protection against electric shock: Class II
Degree of protection against electric shock: Type BF
Equipment not suitable for use in the presence of flammable anaesthetic mixture with air.
*Conditions may vary based on altitude above sea level,barometric pressure, and temperature.
**Value determined with minimum flow rate through the nebulizer (20°C, sea level).
There are no hazardous materials used in the construction of this unit, therefore, there are no known risks
associated with disposal of the equipment.
This product (models 18070) complies with the following electromagnetic compatibility standard:
IEC 60601-1-2:2nd Edition
7.5" W x 4.0" H x 6.8" D (19.1 cm x 10.2 cm x 17.3 cm)
4.1 lbs. (1.86 kg)
120VAC, 60Hz
65 dBA
110 watts maximum
30 psig or greater
10 psig or greater
8 lpm or greater*
+40° to +104°F (+5 to +40°C)
up to 95% non-condensing
up to 95% non-condensing
6 ml (cc)
0.15 ml/min or greater*
5 microns or less**
................................................................................................................
DRIVE PACIFICA ELITE COMPRESSOR/NEBULIZER MODEL 18070
................................................................................................................
13