Amprobe DM-II PLUS Mode D'emploi page 26

Enregistreur de qualité de puissance
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
Message
No parameter se-
lected
No Phase selected
PASSWORD ERROR
PASSWORD OK
Please wait
Recording
Too many param
Too many record
ERR: SEQ
ERR: P-
ERR: SEQ & P-
ERR: CONNECTION
Error Vref
Error1 to Error 5
Description
A recording has been started without
selecting values to be recorded
Voltage and/or current harmonics have
been selected and the corresponding
flag has been enabled (HARMONICS
ON) but no phase voltage or current has
been selected
The password entered is wrong
The password is correct
The instrument is waiting for the record-
ing to start
The instrument is recording
More than 63 parameters have been
selected (harmonics included) or more
than 38 parameters with COGENERA-
TION Flag enabled
The quantity of recorded data + Smp
exceeds the maximum allowed (35)
The Phase Sequence is wrong.
The active powers on the right side of
the message are negative
The active powers on the right side of
the message are negative and the Phase
Sequence is wrong.
The Voltage inputs are not connected
correctly.
A Voltage reference was selected that is
not compatible with voltage input.
The instrument memory is damaged.
24
Advices
Push START/STOP, and se-
lect at least one value from
the MENU
Select at least one phase
voltage and/or current
Insert the password: F1, F4,
F3, F2
Deselect parameters
Transfer recordings to a PC
and then clear the instru-
ment memory
Check the Phase Sequence
connection.
Check if the clamps are
properly connected, unless
co-generation is involved
Check if the clamps are
properly connected / check
the Phase Sequence con-
nection, unless co-genera-
tion is involved
Check the Voltage connec-
tions (see Physical Connec-
tions, Page X8)
Check Voltage Reference
set in RECORDER CONFIG
in MENU
Contact Amprobe for as-
sistance