Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 111
Operating instructions pressure measuring devices SATA adam 2, SATA adam 2 U
Note!
• Further cleaning tips can be found at www.sata.com/TV.
9.

Battery Replacement [5]

• Replace the battery outside of explosive areas only!
• Please observe safety instructions in chapter 6.4!
Only RENATA batteries (CR2032) must be used. Use the battery kit (Art.
No. 160739) consisting of battery, O-rings, locking clip, elastic rubber
strap and special tool. According to the intensity of use, the battery will
last between 1 to 3 years. The battery capacity is electronically controlled.
To exclude measurement errors, the display switches off with insufficient
battery power. In this case, the battery has to be replaced.
Depending on display and use, the battery must be replaced within the
following time frames:
Display:
Battery symbol
Exclamation marks (flashing)
Display "Lo
" upon switching-on
b
1. Disassembly of the display [5-1] (also see chapter 7.1).
2. Open display using the special tool and the elastic rubber strap
[5-2], [5-3].
3. Remove the battery with special tool [5-4].
4. Insert a new RENATA CR 2032 (observe polarity of battery!).
5. Remove the sealing with the special tool [5-5] and insert the new
sealing.
6. Screw the housing back onto the display. To prevent penetration of
cleaning agent, please ensure to tighten the housing hermetically
(completely) [5-3].
7. If required, replace the locking clip and the small O-ring. Flat side of
locking clip must be visible [no illustration].
Warning! Risk of explosion!
4-5 weeks
2-3 weeks
<1 week
EN
81

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Adam 2 miniAdam 2 displayAdam 2 uAdam 2 black

Table des Matières