Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 111
US
Operating instructions pressure measuring devices SATA adam 2, SATA adam 2 U

10. Troubleshooting

Malfunction
No pressure is
displayed
Display cannot
be removed at
all or only with
difficulty
Air leaks be-
tween display
and connection
module dock
Air leaks from
the ball valve [1-
8] (when using
without display)
Display does not
snap-in
Digital display is
either black or
blurred
308
Cause
• Display has not prop-
erly snapped into the
dock
• Battery is empty
• Regulation wing is not
in the horizontal centre
position
• Slide-in groove is
clogged.
• Locking clip has not
snapped in
• Special O-ring [1-6] is
swollen due to soaking
in solvent
O-Ring beneath the lock-
ing clip is missing or dam-
aged
Ball valve is clogged
• Locking clip is dam-
aged.
• Slide-in groove of the
connection module is
clogged or damaged
Display has been exces-
sively exposed to clean-
ing solution
Corrective action
• Slide in the display
until it snaps home,
chapter 7.1
• Replace the battery
• Place regulation wing
in horizontal centre
position, chapter 7.1
• Clean the slide-in
groove.
• Press in the locking
clip, chapter 7.1
• Avoid soaking in sol-
vent, chapter 8
Replace the O-ring,
chapter 9
Clean ball valve, chap-
ter 8
• Replace locking clip,
chapter 9
• Clean the groove
or replace the dock,
respectively
Cleaning, chapter 8,
replacement of the digi-
tal unit

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Adam 2 miniAdam 2 displayAdam 2 uAdam 2 black

Table des Matières