Table 4
Recommended separation distances between portable and mobile RF communications equipment and the SFT53 Ear Thermometer
The SFT53 Ear Thermometer is intended for use in an electromagnetic environment in which radiated therefore disturbances are controlled. The
customer or the user of the SFT53 Ear Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the SFT53 Ear Thermometer as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higer frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
d = [ 3.5
]
E
1
0.12
0.38
1.2
3.8
12
67
(m)
800 MHz to 2.5 GHz
d = [ 7
]
E
1
0.23
0.73
2.3
7.3
23