Safety Designation per IEC 60601-1
Type of protection against
electrical shock
Degree of protection against
electrical shock
Power Equipment
Degree of Electrical connection
between equipment and Patient
Degree of mobility
Mode of operation
Classifications according to IEC 60601-1
MicroLab
Volume Transducer
Option SPO2 Finger Probe
(Nonin IPOD
Integrated Oximetry Device)
®
WARNING: No modification of this equipment is allowed.
Note: When you connect other equipment to the unit, always
make sure that the whole combination complies with the
international safety standard IEC 60601-1 for medical electrical
systems. During measurements, connect the MicroLab only to
printers and computers that comply with IEC 60601-1 / ANSI/AAMI
ES60601-1:2005 / CAN/CSA-C22.2 No. 60601-1:14.
WARNING: The user must not touch any voltage-carrying parts
and the patient at the same time.
During database upload the MicroLab may be connected to a
computer that complies with EN 60950 – 'Information technology
equipment - Safety - Part 1: General requirements'.
Internally powered
Equipment and Class I
Type B applied part
Battery type: NiMH battery
pack, 8.4V, 1100mAh
Equipment designed as
non-electrical connection to
the patient.
Transportable
Continuous
Applied part, type B
Applied part, type B
Applied part, type BF