Cleaning Instructions

Disinfection of contaminated parts is only effective after having
them carefully pre-cleaned. Vyaire recommend the tested solution
of PeraSafe sterilizing powder (36-SSC5000A) for pre-cleaning and
disinfection. If a different solution is used, please follow the given
manufacturer's instructions.
The device must not be wiped with any aqueous solutions and
must not be exposed to solvents e.g. alcohol, chloride solutions as
there are electronic components inside that will be permanently
damaged.
CAUTION: Switch off the device and always unplug your MicroLab
before cleaning.

External Surfaces of the Spirometer

CAUTION: Do not attempt to wash or immerse the MicroLab
transducer housing in water or cleaning fluid, as there are
electronic components inside that will be permanently damaged.
The external housing of the spirometer may be wiped with sterile
wipes or a damp cloth that has been immersed in a cold sterilising
solution, when required.
CAUTION: Do not wipe the touch screen.