Range
200 mA
10 A
Overload protection: F250mA/250V fuse for mA measurement range; F10A/500V
fuse (ceramic) for 10 A measurement range.
Maximum input current: 200 mA DC or AC rms for mA range; 10 A DC or AC rms
for 10 A range.
For measurements > 10 A, the maximum measurement time is 10 s. Do not
measure current for longer than 15 minutes.
Range
20 mA
200 mA
10 A
Overload protection: F250mA/250V fuse for mA measurement range; F10A/500V
fuse (ceramic) for 10 A measurement range.
Maximum input current: 200 mA DC or AC rms for mA range; 10 A DC or AC rms
for 10 A range.
For measurements > 10 A, the maximum measurement time is 10 s. Do not
measure current for longer than 15 minutes.
Frequency response: 40 Hz ~ 400 Hz sine wave rms (average response).
Range
20 kHz
Input voltage range: 200 mV to 10 VAC rms
Overload protection: 250 VDC or 250 VAC rms
Range
200 Ω
2 kΩ
20 kΩ
200 kΩ
2 MΩ
20 MΩ
200 MΩ
Overload protection: 250 VDC or 250 VAC rms
Open circuit voltage: < 700 mV
Range
1 V
Test current: ± 1 mA
V. 03 – 01/02/2013
DVM893
DC current
Resolution
0.1 mA
10 mA
AC current
Resolution
10 µA
0.1 mA
10 mA
Frequency
Resolution
10 Hz
Resistance
Resolution
0.1 Ω
1 Ω
10 Ω
100 Ω
1 kΩ
10 kΩ
0.1 MΩ
Diode
16
Accuracy
± (1.5 % of reading + 1 digits)
± (2.0 % of reading + 5 digits)
Accuracy
± (1.0 % of reading + 5 digits)
± (1.8 % of reading + 5 digits)
± (3.0 % of reading + 7 digits)
Accuracy
± (2.0 % of reading + 5 digits)
Accuracy
± (0.8 % of reading + 3 digits)
± (0.8 % of reading + 2 digits)
± (1.0 % of reading + 2 digits)
± (6.0 % of reading + 10 digits)
Resolution
0.001 V
©Velleman nv