EN-10
4.1.2.1
Auto or Continuous End Point Mode
Press and hold the Auto-Stop button to change the End Point Mode.
When in Continuous Mode, to manually reach a calibration value, you need to
press the Auto-Stop button when reading is stable and displays
reading freezes and
freezes on the display.
When in Auto End Point Mode, the meter determines when the reading is stable
then displays and locks the reading or calibration value automatically, the reading
freezes and
on the display.
4.1.2.2 Calibration
Place the conductivity sensor in the selected calibration standard, wait for 10-15
seconds, then:
1
Press the Cal button.
blinking during calibration.
Note:
a)
Calibration starts with the Auto End Point Mode.
b)
Pressing the Auto-Stop button during the calibration can end the process.
2
The meter reaches endpoint automatically according to the preselected auto-
endpoint mode. The screen will display the calibration K value. The calibrated
standard solution value will be marked in
The calibration is finished, the calibration result will be stored automatically.
Note:
To ensure the most accurate conductivity readings, you should verify your cell constant
with a standard solution before measurement and recalibrate if necessary. Always use
fresh standards.
For STCON3 the normal cell constant range is 1.50 – 2.00. (e.g. 1.71 /cm) if the cell
constant is outside the range due to an improper calibration, you may need to recover
factory settings, then repeat calibration.
4.2

Sample measurement

Place the conductivity sensor in the sample, wait for 10-15 seconds, then:
1
Press the Read/Enter button to start the measurement,
display and is blinking during measurement.
2
When the meter reaches endpoint (
blinks 3 times and freezes on the display.), it will display the conductivity.
blinks 2 times then disappears;
blinks 2 times then disappears;
and
a-AB23EC Bench Conductivity Meter
blinks 3 times and freezes
appear on the top of the screen, and both are
.
blinks 2 times then disappears;
. Then the
blinks 3 times and
appears on the