FREQUENCY OF MAINTENANCE AND INSPECTION OPERATIONS
Perform the following checks on all FENZY BIO-S-CAPE units:
Before use:
•
Check the pressure in the cylinder
•
Check the seal
Periodic annual visual checks:
•
Check the seal
•
Check the pressure gauge
•
Check the external appearance of the apparatus
Every 6 years:
•
Complete service of the FENZY BIO-S-CAPE, including the pressure reducing valve, must be performed by an agent authorised
by HONEYWELL.
If the apparatus is stored under conditions different from those recommended, an annual visual
check of the hood is essential.
SPARE PARTS
The spare parts of the FENZY BIO-S-CAPE device are presented through exploded views in the spare parts price list (price and part
number), which is available on request.
SUPPORT AND TRAINING
All HONEYWELL apparatuses are designed to allow maintenance by the users, but that requires a minimum level of skills and the use of
appropriate equipment.
Training courses may be organised on the customer's premises or in our own training establishments.
The full programme of maintenance training courses for the FENZY BIO-S-CAPE is available on request.
For more information, please contact the HONEYWELL technical assistance department.
LIMIT OF MANUFACTURER'S GUARANTEE
The FENZY BIO-S-CAPE carried the HONEYWELL standard guarantee. In compliance with our own manufacturer's instructions,
inspection and maintenance of the FENZY BIO-S-CAPE must only be carried out by qualified professionals, trained and approved by our
technical department. Use only replacement parts supplied by the manufacturer and always follow the inspection and maintenance rules
specified in these instructions.
EN-5