Self-Diagnosis; Recover Factory Settings - OHAUS STARTER 300 Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

STARTER 300 Portable pH Meter
5.4

Self-diagnosis

Press and hold
icon blinks one after the other.
This checks whether all icons are correctly shown. The next step is to check that the keys are
functioning correctly. This requires user interaction.
When b blinks, five icons are displayed.
Press the five keys in any order. Each time you press a key an icon disappears from the
screen, continue to press the other keys until all the icons have disappeared.
When the self-diagnosis has been completed successfully, PAS appears. If self-diagnosis fails,
error message Err 1 appears. In this case, repeat the self-diagnosis.
Note: You have to finish pressing all five keys within 2 minutes, otherwise Err 1 appears and you
will have to repeat the procedure.
5.5

Recover factory settings

When the meter is off, press and hold
screen displays "RST" and blinks, press
offset, etc.).
Press
to switch off the meter.
and
simultaneously until the meter displays the full screen. Each
&
&
together for 3 seconds, the
to reset factory settings (MTC, slope and
EN-15