EN-8
3
STARTER 300 SETUP
3.1
Set temperature unit and MTC value
Please note:
If a temperature electrode is used, Automatic Temperature Compensation (ATC) and the sample
temperature are displayed on the screen. You may then choose to skip MTC setup (below).
If the meter does not detect a temperature electrode or one is not used, the meter automatically
switches to Manual Temperature Compensation (MTC) mode and MTC appears on the screen.
MTC can be set as follows:
Power the meter on by pressing
Press and hold
temperature unit blinks (°C or °F).
Press
or
Press
to confirm your selection.
Then
Continue with MTC
compensation accordingly
Press button-
Press
to return to the measurement screen.
The default MTC temperature value setting is 25 °C (77°F).
°C = 5/9 (°F - 32)
Note:
3.2
Selecting a predefined buffer group
After confirming the MTC value (or if you are using ATC) you have the option of making a buffer
group selection. In most cases the default buffer group will be used, buffer group b1.
If it is necessary to switch the buffer group, use button-
group among 4 buffer groups shown in section 4.1.1 in this manual. Press button-
the setting or press button-
The default buffer group is b1.
1.68
b1
until the setup icon
to switch between °C and °F .
temperature setting by using
to confirm the setting
to return to the measurement screen.
4.01
7.00
STARTER 300 Portable pH Meter
appears on the display and the current
10.01
or
to adjust temperature
or button-
to select a buffer
to confirm