Assembly
Check to ensure that the following components are included
Mast and lift arm, actuator, slingbar, control box, hand control.
Undercarriage with base-width adjustment motor, screws,
washers, nuts and screw covers.
User manual and charger.
Insert the mast onto the base. Assemble according to illustration A and tighten it
hard (65 ± 5 Nm). Loctite. Place the screw covers over the screws.
Place the battery in the mount in the control box.
Connect the cables: The cable for the hand control in outlet HS; the cable for the
hoist motor in outlet M1, and the cable for the base motor in outlet M2.
Release the emergency stop and perform a final inspection (see final inspection).