Cleaning the Transducer
The transducer requires no routine maintenance or servicing.
However, if you wish to disinfect or clean the transducer it may be
removed by means of the following procedure:
1. Rotating the turbine transducer anti-clockwise until the
locating pip lines up with the small rectangular cut-out in the
housing as shown below.
2. Gently pull the transducer away from the housing.
3. The transducer may now be immersed in warm soapy water
for routine cleaning or immersed in a cold disinfectant solution
e.g. Perasafe (Cat No. 36-SSC5000A) for a maximum of 10
minutes (Alcohol and chloride solutions should be avoided).
4. After cleaning/disinfecting, the transducer should be rinsed
briefly in distilled water and dried.
5. Re-assemble the mouthpiece holder.
Turbine transducer
CAUTION: Do not attempt to wash or immerse the SpiroUSB
transducer housing in water or cleaning fluid, as there are
electronic components inside that will be permanently damaged.
Housing
Locating Pip
Cut-Out
53