Cleaning The Transducer - Vyaire SpiroUSB Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 72

Cleaning the Transducer

The transducer requires no routine maintenance or servicing.
However, if you wish to disinfect or clean the transducer it may be
removed by means of the following procedure:
1. Rotating the turbine transducer anti-clockwise until the
locating pip lines up with the small rectangular cut-out in
the housing as shown below.
2. Gently pull the transducer away from the housing.
3. The transducer may now be immersed in warm soapy water
for routine cleaning or immersed in a cold disinfectant
solution e.g. Perasafe (Cat No. 36-SSC5000A) for a
maximum of 10 minutes (Alcohol and chloride solutions
should be avoided).
4. After cleaning/disinfecting, the transducer should be rinsed
briefly in distilled water and dried.
5. Re-assemble the mouthpiece holder.
Turbine transducer
CAUTION: Do not attempt to wash or immerse the SpiroUSB
transducer housing in water or cleaning fluid, as there are electro-
nic components inside that will be permanently damaged
Housing
Locating Pip
Cut-Out
25

Publicité

Table des Matières
loading

Table des Matières