4. Operation
4.5. EC and NaCl measurements
Measurement by stability
The value is fixed on the display when during 6 seconds the
reading variation is less than 1 digit.
During measurement, the measured value appears on the display
from time to time to orient the user.
Continuous measurement
The instrument displays the value obtained by the electrode all
the time.
10-04-2010
17:15
23.2˚C
TRef 25˚C
New CAL. in 7 d
MEASURING
by stability
23.2˚C
* * *
uS/cm @ 25˚C
Time: 1 s
to vary stir-
Press
ring speed
MEASURING
by stability
23.2˚C
* * *
uS/cm @ 25˚C
Stirring 30 %
MEASURE STABLE
23.2˚C
728
uS/cm @ 25˚C
Time: 15 s
With conductivity cell
without temperature
probe.
MEASURING
by stability
21˚C
* * *
uS/cm @ 25˚C
Time: 1 s
Temperature varies
MEASURING
by stability
19˚C
* * *
uS/cm @ 25˚C
Time: 1 s
Stirring varies.
MEASURING
by stability
19˚C
* * *
uS/cm @ 25˚C
Stirring
30
%
etc.
8
sensION+ EC7 HACH_LANGE 0710.indd ANG:8
sensION+ EC7 HACH_LANGE 0710.indd ANG:8
MEASURING
MEASURING
continuous
by stability
23.2˚C
12.28
* * * * mg/l
mS/cm @ 25˚C
Time: 1 s
Time: 1 s
MEASURING
MEASURING
continuous
by stability
23.2˚C
12.30
* * * * mg/l
mS/cm @ 25˚C
Stirring 30 %
Stirring 30 %
MEASURING
MEASURE STABLE
continuous
23.2˚C
12.30
228 mg/l
mS/cm @ 25˚C
Time: 15 s
Time: 22 s
HOLD
23.2˚C
12.32
mS/cm @ 25˚C
Continuous
measurement.
Time: 17 s
4.6.Other options
TRef and TC selection, change of temp. units
By default the instrument is programmed with TRef 25ºC and TC
2.00 %/ ºC To modify these values, follow the scheme.
10-04-2010
17:15
23.2˚C
TRef 25˚C
New CAL. in 7 d
10-04-2010
17:15
TRef 25˚C
10-04-2010
17:15
TRef 20˚C
10-04-2010
17:15
TC:
2.00
%/˚C
10-04-2010
17:15
TC:
1.90
%/˚C
10-04-2010
17:15
25
˚C
10-04-2010
17:15
25
˚F
10-04-2010
17:15
23.2˚C
TRef 20˚C
New CAL. in 7 d
8/7/10 16:49:57
8/7/10 16:49:57