Télécharger Imprimer la page

Velleman APPA93N Mode D'emploi page 4

Publicité

(4) AC Current
Range
200 µA
20mA
2mA
200mA
10A
AC Conversion Type : Average sensing rms indication.
Overload Protection : 1A/500V fast blow for mA, µA input.
16A/500V fast blow for 10A input.
* Plage 20A : max. 30 secondes sur l'entrée 10 A
(5) Resistance
Range
Resolution
200Ω
0.1Ω
2KΩ
20KΩ
10Ω
200KΩ
100Ω
2MΩ
1KΩ
20MΩ
10KΩ
Overload Protection : 600V d.c/a.c max.
(6) Diode Check
Range
Resolution
1mV
* For 0.4V ~ 0.9V
Overload Protection : 600V d.c./a.c. max.
Continuity : Internal sounder operates when resistance is less than 50Ω .
(7) Auto Power Off
The meter will automatically shut itself off after approximately 30 minutes after power on.
The meter can be turned back on by switching to another range.
Resolution
0.1 µA
1 µA
±(1.5%reading + 3digits)
40Hz — 500Hz
10 µA
100 µA
±(2.5%reading + 7digits)
10mA
40Hz — 500Hz
Accuracy
±(0.8%reading + 4digits)
±(0.8%reading + 1digit)
±(1.5%reading + 5digits)
Accuracy
±(1.5%reading + 5digits)
Accuracy
Voltage Burden
600mV rms max.
900mV rms max.
Max.Test
Current
2.5mA
200 µA
40 µA
4 µA
400nA
40nA
Max.Test
Current
1.5mA
Max.Open
Circuit Voltage
3.2V
0.5V
Max.Open
Circuit voltage
3.2V

Publicité

loading